Stats Digest Feeds
← Back to all CVEs

CVE-2021-4479

MEDIUM NVD
CVSS Score 4
Severity MEDIUM
Published Jun 02, 2026
Vendor unknown

Description

Dräger Atlan A350 software versions 1.00 through 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.

References